Accuracy and Precision of Three Acceleromyographs, Three Electromyographs, and a Mechanomyograph Measuring the Train-of-four Ratio in the Absence of Neuromuscular Blocking Drugs
Wedemeyer Z, Michaelsen KE, Jelacic S, et al. · Anesthesiology 2024;141(2):262-271
In 28 patients given no neuromuscular blocking drugs, 9,498 train-of-four ratios were collected from three acceleromyographs (Stimpod™ among them), three electromyographs (Stimpod™ among them) and a laboratory mechanomyograph. The Stimpod™ acceleromyograph recorded the lowest overshoot, the lowest patient-to-patient variability and the tightest range of the acceleromyographs tested. The three electromyographs, including the Stimpod™ electromyograph, matched mechanomyography closely (estimated means 0.99 to 1.01; gauge SDs 0.01 to 0.02).